[J48] High Energy X-ray Imaging System Using Silicon Strip Detector

XIAOSONG YAN1, HIROYUKI TAKAHASHI1, KENJI SHIMAZOE1, TAKESHI FUJIWARA1, TAKESHI ISHIKURA1, YANG TIAN1 (1.Department of Nuclear Engineering and Management, The University of Tokyo)

The imaging system is designed for disaster recovery and prevention by using ~5 MeV X-ray. We designed the system to be 4 x 4 modules, and one module constitutes with 20 stacked detectors with tungsten collimators and ASIC. We have tested the silicon strip detector, the strip detector has ~7 keV FWHM at 60 keV when a bias of 120 V is applied. We also tested the detector with ASICs, and the position identification has been confirmed.
In the future, we will continue to test with accelerator X-ray beam, and optimize the configuration of the system.