3:30 PM - 3:45 PM
[MAG37-01] Operation of the Tokai Radionuclide Laboratory
Keywords:CTBT, IMS, Tokai Radionuclide Labolatory
In January 2020, we replaced the RL11’s analysis system due to the progressive deterioration of the high-purity germanium semiconductor detector over time. The technical issue with implementation of the new analytical system was to determine the detection efficiency of six different shaped samples sent from the radionuclide monitoring station. We calculated the detection efficiency based on the geometry of the sample using software, and adjusted the geometry to be more accurate based on the results of measurements using radionuclide sources with known activity. In addition, we introduced an end cap with a copper plate to prevent errors in the calculation of activity due to X-ray interference.
The CTBTO conducts an annual Proficiency Test Exercises (PTE) as part of the quality assurance of analyses at the radionuclide laboratories. At the PTE, we perform detailed analyses of unknown samples prepared for the PTE and report the results to the CTBTO. April 2021, we participated in PTE2021 with the new analysis system and passed all the assessment points and received the highest rating.
In this presentation, I will report the activities of the Tokai Radionuclide Laboratory (RL11) in the CTBT verification regime and the introduction of the new analytical system.