[PCG23-09] 火星探査におけるネオン測定に向けた分別膜の性能評価

*奥野 衛1吉岡 和夫1三浦 弥生2長 勇一郎3斎藤 義文4笠原 慧1杉田 精司1 (1.東京大学大学院理学系研究科地球惑星科学専攻、2.東京大学地震研究所、3.NASA Marshall Space Flight Center、4.宇宙航空研究開発機構・宇宙科学研究所)

キーワード:Martian atmosphere, Ne measurement, Mars mission, mass spectrometer, instrument development

Present Mars has cold and dry climate with a very thin atmosphere. However, early Mars may have possessed warm and wet climate with a large amount of atmosphere. One possible cause for such drastic change in climate and atmospheric mass on Mars is escape of atmosphere to space, but actual process of such possible atmospheric loss has not been understood well yet. One of the reasons for this uncertainty comes from the lack of our knowledge on how much Martian atmosphere has been lost. Noble gases, which are chemically inactive, are important for estimating the degree of the atmospheric loss. The non-thermal escape, such as pick-up ion sputtering, induces isotopic fractionation because lighter isotopes are selectively lost, resulting isotopic compositions in the atmosphere to be heavier. Since light noble gases, such as Ne, are sensitive to such a mass fractionation process, its isotopic ratios are useful for constraining the degree of the atmospheric loss. Thus, isotopic ratios of Ne are important in understanding the evolution of Martian atmosphere.
However, the Ne isotopic composition of Martian atmosphere has not been measured directly either by Viking or Curiosity. A typical lightweight mass spectrometer with moderate mass resolution (m/Δm ~ 100), such as a quadrupole mass spectrometer (QMS), cannot distinguish 40Ar++ from 20Ne+ signal because the difference in their mass/charge ratios is very small (m/Δm = 1777). Thus, Ar needs to be removed from sample gas before mass spectrometric measurements.
In order to achieve direct measurements of the Ne isotopic ratio with a future Mars lander, we are developing a gas measurement system with a permeable membrane. In this study, we experimentally investigate the difference in permeability of Viton sheets between Ar and Ne. Our experimental results indicate that a Viton sheet with 1 mm in thickness can increase the abundance ratio of Ne to Ar from the atmospheric value of ~10-3 to ~1. We also measured the ratio of 40Ar++ to 40Ar+ using a QMS with a typical ionization voltage (70V). The results show that the amount of 40Ar++ produced during the ionization process in the QMS is about 10% of that of 40Ar+. Thus, gas sample permeated through the Viton sheet would have 40Ar++ contribution about 10% of 20Ne+ contribution. These results suggest that the Ne isotope measurements can be achieved with uncertainty better than 10% after correcting for the contribution of 40Ar++. Since Martian atmospheric pressure is about two orders of magnitude lower and 20Ne/40Ar ratio is about one order of magnitude lower in Martian atmosphere, the separation efficiency could decrease compared to that at the terrestrial atmospheric condition. However, it has a room for significant improvement by optimization for various parameters, such as materials, thickness, and duration of permeation. These results suggest that measurements of Ne isotopic ratio in Martian atmosphere may be achieved with this approach after optimization.