10:30 AM - 12:10 PM
[3Rin2-47] Evaluation of Automatic Monitoring of Instillation Adherence Using Eye Dropper Bottle Sensor and Deep Learning in Patients with Glaucoma
Keywords: Medication information, IoT, Deep Learning
Purpose: We developed and evaluated an eye dropper bottle sensor system comprising motion sensor with automatic motion waveform analysis using deep learning (DL) to accurately measure adherence of patients with antiglaucoma ophthalmic solution therapy. Results: The developed eye bottle sensor detected all 60 instillation events (100%). Mean (SD) difference between patient and eye bottle sensor recorded time was 1 (1.22) (range; 0–3) min. Additionally, mean (SD) instillation movement duration was 16.1 (14.4) (range; 4–43) s. Two-way ANOVA revealed a significant difference in instillation movement duration among patients (P<0.001) and across days (P<0.001). Conclusion: The eye dropper bottle sensor system developed by us can be used for automatic monitoring of instillation adherence in patients with glaucoma.