11:00 AM - 11:15 AM
△ [15a-2W-8] Large-dose and high-resolution X-ray dosimeters for microbeam radiation therapy
Keywords:Dosimeter,Samarium,Microbeam Radiation Therapy
Microbeam Radiation Therapy (MRT) is a potential canther therapy using intese and high-energy synchrotron X-rays. The measurement of X-ray dose distirbution in MRT requires both dynamic range of 5-1000 Gy and a-few-micrometer resolution together; however, no existing detectors have been able to handle such high demanding requirements. In this study, we have developed a state-of-the-art technique using Sm-doped radiophotoluminescence (RPL) dosimeters, which relies on the Sm3+-to-Sm2+ interconversion associated with X-ray irradiation. With our new technique, we have successfully demonstrated accurate measurements of X-ray dose distribution in MRT.