[19a-W810-10] Spectroscopic Measurements of L X rays emitted from 244Cm sources with using TES microcalorimeters

Keisuke Maehata1, Tetsuya Sugimoto1, Naoko Iyomoto1, Kenji Ishibashi1, Keisuke Nakamura2, Yuki Morishita2, Koji Takasaki2, Kazuhisa Mitsuda3 (1.Kyushu Univ., 2.JAEA, 3.JAXA)

Keywords:TES microcalorimeter,Transuranium elements,Spectroscopic measurement of L X ray

L X rays with the energy from 10 keV to 20 keV are emitted from most transuranium elements (TRU) following their α decays with relatively high probability. 244Cm is a typical TRU in a nuclear fuel reprocessing. In this work, spectroscopic measurement was conducted for L X rays emitted from a 244Cm source with using a TES microcalorimeter. The TES microcalorimeter consisted of a Au absorber of 5 μm thick. The surface area of the Au absorber was 160 x 160 μm2. The energy resolution of the TES microcalorimeter was obtained to be 39 eV FWHM by analyzing a Pu Lα1 peak of 14.28 keV in the experimental energy spectrum of L X rays emitted from the 244Cm source.