5:45 PM - 6:00 PM
[21p-S422-15] Multi-wavelength Swept Interferometric Microscope for en-face Tomographic Vibration Measurement in Biological Tissue
Keywords:En-face optical coherence tomography,Biological vibration measurement,Mechanobiology
To address microvibrations occur in bio-tissues, multi-wavelength swept OCT utilizing wide-field heterodyne vibrometry has been developed. These arrangements allow obtaining not only 3D tomographic images but also surface vibration parameters such as spatial amplitude, phase, and frequency, over a wide field. Our system achieved the axial resolution of ~2.5 μm with measurable depth of 0.5 mm, and we captured vibrations of internal surfaces of bio-tissue at frequency of 1 – 10 kHz