3:45 PM - 4:00 PM
△
[5p-S43-10] Fabrication of ComptonPET hybrid camera using Ce:Gd3(Ga,Al)5O12 and Eu:SrI2 and
test of simultaneous imaging of multi-RI tracers using 18F and 111In
Keywords:Scintillators, Positron Emission Tomography, Compton camera
An array detector was prototyped by combining MPPC array with Ce:GGAG and Eu:SrI2, and the energy resolution and the linearity of the output response with respect to gamma ray energies were evaluated. As an example, an energy resolution of 15.5% ± 1.1% FWHM was obtained with 57Co at 122 keV gamma-rays. Performed test of multi-tracer simultaneous imaging using 18F and ¹¹¹In was demonstrated, and it is possible to image 18F and ¹¹¹In independently. In this presentation, we also report on the evaluation results of Eu:SrI2 array detector, and discuss the characteristics required for the scintillator to use for ComptonPET application.