The 78th JSAP Autumn Meeting, 2017

Presentation information

Oral presentation

3 Optics and Photonics » 3.8 Optical measurement, instrumentation, and sensor

[6p-A414-1~21] 3.8 Optical measurement, instrumentation, and sensor

Wed. Sep 6, 2017 1:15 PM - 7:00 PM A414 (414)

Tatsutoshi Shioda(Saitama Univ.), Takeshi Yasui(Tokushima Univ.), Yu Tokizane(RIKEN)

3:30 PM - 3:45 PM

[6p-A414-9] Optical-frequency-comb thermometer II

Yukiko Shimizu1,2, Sho Okubo1,2, Atsushi Onae1,2, Hajime Inaba1,2 (1.NMIJ, AIST, 2.JST, ERATO MINOSHIMA IOS)

Keywords:dual-comb spectroscopy, temperature, gas thermometry

Acetylene gas temperature is measured using dual-comb spectroscopy. We found that the fitting of absorption profile of many rotational spectral lines of the specific vibrational transitions to a theoretical equation give a precise temperature. The spectroscopic temperature is determined with the estimated uncertainty of 1K at 23.6 ℃.