3:30 PM - 3:45 PM
[6p-A502-8] Sensitivity enhancement of V-trench biosensors by time-resolved imaging
Keywords:optical biosensor, fluoroimmunoassay, quantum dot fluorescent dye
The V-trench biosensor is a miniature and sensitive sensor for fluoroimmunoassay. For further sensitivity, autofluorescence of the sensor chip substrate need to be totally removed. To this end, time-resolved imaging of the V-trench biosensor was performed utilizing the difference of fluorescence lifetime between luminescence signals and autofluorescence. A time-resolved imaging system equipped with a fast-gated image intensifier was developed, and the sensitivity was improved by 4-fold compared to the measurements without the time-resolved method. Moreover, the measurements using the CMOS image sensor for fluorescence lifetime imaging were performed.