The 64th JSAP Spring Meeting, 2017

Presentation information

Oral presentation

2 Ionizing Radiation » 2.2 Detection systems

[15a-E204-1~5] 2.2 Detection systems

Wed. Mar 15, 2017 10:30 AM - 11:45 AM E204 (E204)

Kenichi Watanabe(Nagoya Univ.)

11:15 AM - 11:30 AM

[15a-E204-4] Measurements and Application of Li-glass Scintillator by the Current Mode

Tetsuro Matsumoto1, Akihiko Masuda1, Hideki Harano1, Jun-ichi Hori2, Tadafumi Sano2 (1.NMIJ/AIST, 2.Kyoto Univ.)

Keywords:BNCT, Li-glass scintillator, current mode

We have developed a pair neutron detector of 6Li-glass and 7Li-glass scintillators in order to measure high-flux neutrons in a BNCT facility. The 7Li-glass scintillator is used to evaluate gamma rays detected by the 6Li-glass scintillator. As a starting point of characterization, we measured the dark current of photo-multipliers, 6Li content in the 7Li-glass scintillator and the linearity of counts from the detector for the neutron flux.