The 70th JSAP Spring Meeting 2023

Presentation information

Poster presentation

2 Ionizing Radiation » 2 Ionizing Radiation (Poster)

[17p-PA09-1~76] 2 Ionizing Radiation (Poster)

Fri. Mar 17, 2023 4:00 PM - 6:00 PM PA09 (Poster)

4:00 PM - 6:00 PM

[17p-PA09-66] Examination of proton beam dose distribution measurements using OSL characteristics of BeO ceramics

〇(M1)Kazuki Omi1, Go Okada2, Weishan Chang1, Yusuke koba3, Ema Sasaki1, Yuta Hirai1, Mitsuki Kawane1, Hinata Huziwara1, Ayaka Yamazaki1, Kenichi Watanabe4, Kiyomitsu Shinsho1 (1.Tokyo Metropolitan Univ Radi., 2.Kanazawa Inst of Technology., 3.NIRS., 4.Kyusyu Univ.)

Keywords:OSLD, proton therapy, beryllium oxide

We investigated the measurement of proton beam dose distribution by our original OSL distribution measuring instrument using a 10.4 cm square BeO ceramic. Dose response and deep dose distribution for proton beam were measured using HIMAC. The dose response characteristics showed good linearity. The deep dose distribution was obtained at intervals of about 70 μm. On the other hand, optimization of sensitivity inhomogeneity and background subtraction was not sufficient.