[14a-P2-6] Optical feedback in a distributed feedback quantum cascade laser and linewidth reduction for cavity ring-down spectroscopy

〇(PC)Volker Thomas Sonnenschein1、Ryohei Terabasyashi1、Noriyoshi Hayashi1、Shuusuke Kato1、Hideki Tomita1、Lei Jin1、Masahito Yamanaka1、Norihiko Nishizawa1、Atsushi Sato2、Kohei Nozawa2、Kenta Hashizume2、Toshinari Oh-Hara2、Tetsuo Iguchi1 (1.Nagoya Univ.、2.Sekisui Med. ADME&Tox.)

キーワード:QCL, feedback, linewidth

Maximal transmission through a high finesse optical cavity, such as those used in Cavity Ring-Down Spectroscopy, is only achieved if the linewidth of the exciting laser is at least as narrow as the transmission peak given by the Airy function. For our system a linewidth of less than 50kHz would be required, which is not generally achieved using standard commercial distributed feedback (DFB) or external cavity QCL lasers, which typically feature a linewidth of more than 1MHz.
Optical feedback is known to reduce the short-term linewidth of lasers. We present results based on a simple external feedback mirror, demonstrating the effects on tuning behavior and linewidth narrowing.