9:30 AM - 11:30 AM
[15a-P1-5] New control method of gradient coil for rotary magnet field cycle MRI
Keywords:MRI, magnetic resonance
We develop new type field cycle MRI using rotary magnet. This system is able to change main magnetic field during MRI imaging. Therefore it gets a lot of attention as elemental technology for DNP-MRI.A problem of rotary magnet MRI is magnetic field distortion at phantom position. Because of distortion, MRI image has serious artifacts.We report the reason of distortion and the control method.