[19p-W810-1] A Scintillation X-ray Detector Using Proportional-mode Si-APD and HfO2 – loaded Plastic Scintillator

Keisuke Inoue1, Shunji Kishimoto2, Rie Haruki2, Masanori Koshimizu3, Tomoyuki Hirata3, Fumihiko Nishikido4 (1.SOKENDAI, 2.KEK / IMSS, 3.Tohoku Univ., 4.NIRS)

Keywords:X-ray Detector,Fast response,Avalanche photodiode

To obtain both a fast response and a detection efficiency sufficient for high-energy X-rays of >30 keV, we tested a scintillation detector using a proportional-mode silicon avalanche photodiode. In case of using a HfO2-loaded plastic scintillator, a good time resolution of 0.34 ±0.06ns (FWHM) was obtained for 57.6 keV X-rays. This was a better value than that of a Pb-loaded plastic scintillator (0.54±0.06 ns).