9:30 AM - 9:45 AM
[18a-C212-3] Measurement of carbon concentration in silicon crystal
(XIX) 2nd generation infrared absorption standard measurement procedure
Keywords:CZ Silicon crystal, carbon concentration measurement, infrared absorption
Standard measurement procedure for the infrared absorption measurement of low concentration carbon was established, both for single and poly crystals, and detection limit 1x1013 and quantification limit 1x1014 atoms/cm3. Te procedure consists of (1) preparation of synthetic reference sample, (2) aquizition of absorption spectra from both measurement and reference samples, (3) processing of differential absorption spectrum, (i) long (573-618-637 cm-1 three point) and short (592-618 cm-1) baselines (ii) deletion of unwanted phonon fractional absorption at 600, 608 and 612 cm-1. (4) determination of carbon peak absorbance.