4:15 PM - 4:30 PM
[MTT38-04] Characterization of MEMS atmospheric pressure sensors for infrasound observation utilization
Keywords:Barometric pressure sensors, pressure gauge calibration, temperature characteristics
For the evaluation of temperature and humidity characteristics, MEMS sensors were placed in a constant temperature and humidity chamber, and the temperature and humidity inside the chamber were changed step by step. The output of the sensors during the temperature and humidity change was compared with the output of a reference atmospheric pressure gauge placed outside the chamber (at room temperature). Temperature was varied from -20 °C to 50 °C while keeping humidity constant to 50 %, and humidity was varied from 40% to 90% while keeping temperature constant to 23 °C. As a result, it was found that the absolute values of temperature coefficients ranged from 0.1 Pa/K to approximately 7 Pa/K, indicating significant variations in temperature characteristics depending on the model. It was also observed that the output of some models was influenced by humidity. For commonly used models, individual differences in temperature and humidity characteristics were examined revealing generally consistent temperature characteristics but some variation in humidity characteristics.
Next, for the evaluation of sensor responsiveness, MEMS sensors were fixed to a table of a vertical electric actuator and subjected to atmospheric pressure fluctuations by moving the table up and down to confirm responsiveness. A height variation of approximately 87 mm corresponds to a pressure change of approximately 1 Pa in the laboratory environment. Results showed that when fluctuations ranging from 1 Pa to 10 Pa were applied at a frequency of 1 Hz, some sensors could detect even a 1 Pa pressure change, while others were unable to detect a 1 Pa pressure change buried within the output noise or produced large output values in response to input. It is important to note the various characteristics when precise atmospheric pressure observations are required. This presentation will report on the details of the evaluation.