[1E08-13] Dosietry 3

Mon. Mar 26, 2018 2:45 PM - 4:20 PM Room E (R1-311 -R1 Building)

Chair: Jun Goto (Niigata Univ.)

3:45 PM - 4:00 PM

[1E12] Determination of the conversion coefficients from air kerma to dose equivalents at the depth of 3 mm for X-ray reference fields

*Masahiro Kato1, Tadahiro Kurosawa1, Hiroyuki Ohguchi2, Kazufumi Yachi3,2,1 (1. AIST, 2. CHIYODA TECHNOL, 3. Hitachi High-Technologies)

Keywords:Radiation protection, eye lens, equivalent, calibration

The conversion coefficients from air kerma to dose equivalents at the depth of 3 mm, H(3), were derived in order to enable calibration and testing in units of H(3) in the X-ray reference fields specified in ISO4037-1, JIS Z 4511 and IEC 61267. The conversion coefficients were obtained based on the conversion coefficients of mono-energetic photons obtained by a simulation method and the energy spectrum with measurements or calculations.